61 Using Setup Mode GX-2003 Operator’s Manual
2. Use the AIR or (SHIFT) button to display the desired
character, then press and release the POWER/ENTER button to
enter the character. The next character flashes.
3. Repeat step 2 to enter the remaining characters. The serial number
is saved and the main menu displays after you enter the last
character.
Turning the User/Station ID Function On or Off
With USER/STATION ID set to ON, the ID Screen displays during
start up and in Display Mode. The ID’s can be updated in Display
Mode.
With USER/STATION ID set to OFF (factory setting), the ID Screen
does not display during start up or in Display Mode.
1. From the main menu, place the cursor next to the USER/
STATION ID menu item. Press and release the POWER ENTER
button.
2. Use the AIR or (SHIFT) button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Calibrating Using Auto Calibration
If the CAL. EXPIRED ACTION is set to CANNOT USE and
calibration becomes due, then the unit must be calibrated using this
menu item and/or the next menu item, SINGLE CALIBRATION
because it will not be possible to enter normal operation. See
“Calibration Supplies and Equipment” on page 32 for the items you
will need to perform a calibration.
1. Install the demand flow regulator onto the calibration cylinder.
2. Connect the sample tubing to the demand flow regulator.
SET
378010028
21:09
SERIAL NO.
SET USER/STATION ID
INPUT
OFF
21:09