OPERATING INSTRUCTION MANUAL
FOR
SMART GAS DETECTOR/TRANSMITTER
MODEL SD-705EC
FOR USERS
1. Read and understand the instructions in this manual before operating this instrument.
2. Keep manual accessible at all times.
3. This instrument cannot be used for any other purpose than what is specified in this
manual.
4. Follow all the instructions in this manual, an deviation will compromise the safety, quality
and performance of this instrument.
5. We accept no responsibility for an accident caused by
t
he user not following the
instructions in this manual.
Safety Precautions
Phone : 81-3-3966-1113
Fax : 81-3-3558-9110 GIII
2-7-6 Azusawa Itabashi-ku Tokyo, 174-8744 Japan
Seitenansicht 0
1 2 3 4 5 6 ... 37 38

Inhaltsverzeichnis

Seite 1 - FOR USERS

PT2E-1140 OPERATING INSTRUCTION MANUAL FOR SMART GAS DETECTOR/TRANSMITTER MODEL SD-705EC FOR USERS1. Read and understand the instructio

Seite 2

PT2E-1140 2−5 Installation Method (1) Install the detector body to a firm surface (wall surface, etc.) with M6 bolts. Use an optional mounting piec

Seite 3 - Contents

PT2E-1140 (b) Installation using a drip-proof cover(option) When installing the cover, slide it from the top to downward along the groove and fix it w

Seite 4 - 1.Function of the Product

PT2E-1140 (2) Insert a packing gland(lower) →washer→packing→packing grand(upper) in this order onto the cable. Lead the cable into the dete

Seite 5 - Display Part

PT2E-1140 ・For connection of terminal plate <In case of direct connection> Peel length of cable end:7 mm(3p terminal), 7mm(6p terminal)

Seite 6 - Internal View

PT2E-1140 2−6 Wiring Method (1) After installation of detector, confirm that there is no error in installation, and carry out connection between equi

Seite 7 - 2.Handling

PT2E-1140 ・Be sure to make ground as the instrument is flame proof design. ・Make ground with ground resistance is below 100Ω ・Be sure not to connect

Seite 8

PT2E-1140 Normally-closed contact(Break contact) at non-existing condition may change to open contact in a moment due to physical shock. Whenever a

Seite 9 - 2−4 Maintenance Space

PT2E-1140 3.Operation 3−1 Startup Method (1) Confirm that the power is not supplied to the unit and removes 6 hexagonal socket headed bolts and lid

Seite 10 - 2−5 Installation Method

PT2E-1140 3−4 Gas Alarm Function When gas concentration exceeds preset alarm level (alarm point), alarm relay contact and alarm light activates. G

Seite 11 - PT2E-1140

PT2E-1140 4.Maintenance and InspectionThe gas detection alarm is kept in continuous operation over a long period of time and must perform a vital rol

Seite 12

PT2E-1140 Introduction Thank you for purchasing our fixed type gas detector/transmitter Model SD-705EC. This is a gas detector to detect toxic gases

Seite 13

PT2E-1140 Contents of inspection Inspection point/item Inspection by every 6 months Yearly inspection Judgment Sensor sensitivity Calibration with gas

Seite 14 - 2−6 Wiring Method

PT2E-1140 LCD indication Menu Displays set potential Zero adjustment mode Span adjustment mode 4mA adjustment mode Alarm point set mode Alarm t

Seite 15 - CAUTION

PT2E-1140 (1) Zero adjustment method ① Confirm with a portable gas detector that the atmosphere around the detector and measuring gas inlet is clean a

Seite 16

PT2E-1140 ⑤ Operate the pump and adjust flowmeter at 1.5l/min. ⑥ When the calibration gas is introduced into the sensor, the indic

Seite 17 - 3.Operation

PT2E-1140 4−4 4mASignal Output Adjustment Method ① Press SET switch● in MENU display ② When signal output adjustment mode is enter

Seite 18 - 3−5 Trouble Alarm Function

PT2E-1140 4−6 Alarm (Transmission) Test Method Alarm functions can be confirmed. When make alarm test (transmission test), announce it to resp

Seite 19 - 4.Maintenance and Inspection

PT2E-1140 4−7 Point Skip Set Method Maintenance mode can be set compulsorily 4∼20mA :2.5 mA(Fix) Alarm contact:OFF ① Press SET switch

Seite 20 - 4−2 Maintenance Mode

PT2E-1140 ・ When open the lid, wait for wore than 30 seconds after put off the power supply. If not, explosion proof performance can no

Seite 21

PT2E-1140 4−10 Measures for Storage or Long-time Shutdown (1) Store the sensor as attached to the detector in a place not exposed to dust and wate

Seite 22 - REMARKS

PT2E-1140 5.Abnormalities and Countermeasures 5−1 Trouble Indication and Countermeasure This section describes a procedure to determine the fault

Seite 23 - Gas bag for exhaust

PT2E-1140 Contents 1.Functions of‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 3 2.Handling 2ー1 Before Initial Use‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 6 2−2 Cautions for installa

Seite 24

PT2E-1140 5−2 Troubleshooting ① PW/TR light is not ON PW/TR light OFF Yes No Confirm the power supply and reset the normal state.Supply voltage DC24

Seite 25

PT2E-1140 ② Zero adjustment impossible Zero adjustment impossible YesYes Zero adjustment mode with fresh air supplied? No Carry out zero adjustment

Seite 26 - 4−7 Point Skip Set Method

PT2E-1140 ③ Span adjustment impossible Span adjustment impossible Yes Calibration gas prepared correctly? No Use the correctly prepared gas for re-

Seite 27 - 4−9 Fuse Change Method

PT2E-1140 6. Definition Electrochemical method This is a principle for the sensor integrated into this detector. For the details, refer to 8

Seite 28

PT2E-1140 7.Scrap of Sensor Scrap of products ・ For used up sensors, be sure to return them to the manufacturer. The return is requested

Seite 29

PT2E-1140 8.Spesifications 8−1 Specifications Type : SD−705EC Detection principle : Electrochemical Cell method Gas to be detected : Toxic gas D

Seite 30 - 5−2 Troubleshooting

PT2E-1140 Power supply : Supply voltage:DC24V±10% Power consumption : Max 2.5W Operating temperature/humidity : -10 ∼ +40℃ 30∼80%RH(no-condensing)

Seite 31 - ② Zero adjustment impossible

PT2E-1140 8−2 Detection Principle The detection principle of this gas detector is electrochemical method. The sample gas is electrolyzed by the ele

Seite 32 - ③ Span adjustment impossible

PT2E-1140 37 9. Warranty RIKEN KEIKI STANDARD WARRANTY GAS DETECTION INSTRUMENTS RIKEN KEIKI CO., LTD. warrants gas alarm equipment manufac

Seite 33 - 6. Definition

PT2E-1140 1.Function of the Product ◇ Overall view and name of each part 〔SD-705EC〕 △警告“強力磁石” △WARNING:MAGNET MODEL TC-7 RIKEN KEIK

Seite 34 - 7.Scrap of Sensor

PT2E-1140 Display Part ① MAINTENANCE/ESC switch‥‥Used for entering into the maintenance mode with the control key. And us

Seite 35 - 8.Spesifications

PT2E-1140 Internal View ①Terminal plate………………… Connected to power source and 4-20mA signal output. ②Fuse…………………………… Fuse for power(0.5A) ③Rely output

Seite 36

PT2E-1140 2.Handling 2−1 Before Initial Use On detecting a toxic gas leakage, this unit show the gas concentration on the LCD and outputs the gas

Seite 37 - 8−2 Detection Principle

PT2E-1140 Be sure to use an optional drip-proof cover when the detector is to be installed out of doors. Caution ! ⑤ Place where the tempe

Seite 38 - GAS DETECTION INSTRUMENTS

PT2E-1140 ② Power cable noise Following is available to reduce the influence of electromagnetic induction noise and electrostatic induction noise from