www.rkiinstruments.com Operator’s ManualSM-2001U & SM-2003U Single Module Calibration StationsPart Number: 71-0130RKRevision: BReleased: 6/17/10

Seite 2 - Warranty

10 • Back PanelFuse & Power Cord PlugA fuse holder is located to the right of the power switch. The fuse holder consists of a panel mounting sock

Seite 3 - Table of Contents

Control Panel, Status LEDs & USB Port • 11Control Panel, Status LEDs & USB PortThe control panel and status LEDs are the same for both types

Seite 4 - 4 • Table of Contents

12 • Control Panel, Status LEDs & USB PortControl Buttons & LEDsFour control buttons are located to the right of the MODE switch. From left t

Seite 5

SM-2001U Instrument Connections • 13Status LEDsTwo status LEDs are located on the top front of the calibration stations. The CAL. status LED is on t

Seite 6 - System Requirements

14 • SM-2003U Instrument ConnectionsSM-2003U Instrument ConnectionsThe GX-2003 fits into the SM-2003U as shown in Figure 4 below. A coiled black tube

Seite 7 - Specifications

SM-2003U Instrument Connections • 15NOTE: The SM-2003U is shipped with the green tube disassembled from the calibration station. See “Parts Assembly”

Seite 8 - About this Manual

16 • OverviewChapter 3: Hardware Setup and Software InstallationOverviewThere are two parts to preparing the SM-2001U and SM-2003U for use: hardware

Seite 9 - Back Panel

Hardware Setup • 17Connecting an Exhaust TubeThe top exhaust fitting port is used to route exhaust away from the calibration station. Install the 5/1

Seite 10 - Fuse & Power Cord Plug

18 • Hardware SetupTo connect calibration gas to the SM-2001U or SM-2003U do the following: 1. If the area around the calibration station is not cons

Seite 11 - MODE Rotary Switch

Hardware Setup • 19that calibration gas is applied during a bump test, the shorter the bump test time, the larger bump test tolerance used to determi

Seite 12 - Control Buttons & LEDs

2 • WarrantyWarrantyRKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performa

Seite 13 - USB Port

20 • Hardware SetupThe parameter settings for each MODE switch position are shown in Table 4 below.Set the MODE switch to the position that correspon

Seite 14 - SM- 2001U

Installing the Single Module Data Viewer Software • 21Installing the Single Module Data Viewer Software 1. Launch Windows®. 2. Exit from all applicat

Seite 15 - SM -2003U

22 • OverviewChapter 4: Using the SM-2001U & SM-2003UOverviewThe SM-2001U and SM-2003U calibration stations are capable of performing bump tests

Seite 16 - Hardware Setup

Bump Testing Instruments • 23the inlet fitting, and then connect the plastic fitting on the end of the straight green tube to the exhaust fitting.If

Seite 17 - Connecting Calibration Gas

24 • Bump Testing Instruments• Apply fresh air to the instrument for 30 seconds.• Perform a zero operation on the instrument.NOTE: If one or more of

Seite 18 - Setting the MODE Switch

Calibrating Instruments • 25CAUTION: When using the GX-2001 with the SM-2001U or the GX-2003 with the SM-2003U, do not turn off the instrument using

Seite 19 - Hardware Setup • 19

26 • Calibrating Instruments 4. Turn on the instrument.• A GX-2003 will begin its normal startup sequence and after a few seconds, the pump will turn

Seite 20

Calibrating Instruments • 27NOTE: If one or more of the sensors fails the zero operation, then the calibration station will abort the calibration and

Seite 21 - Viewer Software

28 • Bump Testing and Calibrating a GX-2003 with 100% Volume MethaneBump Testing and Calibrating a GX-2003 with 100% Volume MethaneThe GX-2003 can ha

Seite 22 - Bump Testing Instruments

Bump Testing and Calibrating a GX-2003 with 100% Volume Methane • 29the back of the station. The CHARGE status LED will begin to blink green. 4. Put

Seite 23

Table of Contents • 3Table of ContentsWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Seite 24 - 24 • Bump Testing Instruments

30 • Bump Testing and Calibrating a GX-2003 with 100% Volume Methanefollowing functions:• It will apply fresh air to the instrument for 30 seconds.•

Seite 25 - Calibrating Instruments

Charging an Instrument in a Calibration Station • 31CAUTION: When using the GX-2001 with the SM-2001U or the GX-2003 with the SM-2003U, do not turn o

Seite 26

32 • Charging an Instrument in a Calibration Station 4. If the battery pack is fully charged, then the CHARGE status LED will turn solid green in a f

Seite 27 - Calibrating Instruments • 27

Calibration and Bump Test Records • 33Calibration and Bump Test RecordsThe SM-2001U and SM-2003U save a record of each bump test and calibration perf

Seite 28

34 • Calibration and Bump Test RecordsCopying Calibration and Bump Test Records to a USB Flash DriveThe calibration station will only perform a copy

Seite 29

Calibration and Bump Test Records • 35 4. Wait until the flash drive’s LED stops blinking, then remove the flash drive from the USB port.NOTE: If you

Seite 30

36 • Calibration and Bump Test RecordsBump Testing or Calibrating and Saving Files To a Flash Drive Multiple Times In One DayThe SM-2001U and SM-2003

Seite 31 - Calibration Station

Overview • 37Chapter 5: Single Module Data Viewer SoftwareOverviewThe Single Module Data Viewer Software is used to view, organize, and print bump te

Seite 32

38 • Using the Single Module Data Viewer SoftwareNOTE: When you start the Single Module Data Viewer program for the first time, there will be no data

Seite 33

Using the Single Module Data Viewer Software • 39upper right side of the window shows the contents of the item selected in the upper left side, and t

Seite 34 - Flash Drive

4 • Table of ContentsCalibration and Bump Test Records . . . . . . . . . . . . . . . . . . . . . . . . 33Chapter 5: Single Module Data Viewer Softw

Seite 35

40 • Using the Single Module Data Viewer Software 2. Navigate to the location of the files that you want to import into the database and select them.

Seite 36

Using the Single Module Data Viewer Software • 41 2. ID View FormatIn ID view format, the data can be organized by one or more of the following items

Seite 37

Using the Single Module Data Viewer Software • 42Once you have selected how you want to organize the data: 1. Click the expanded view symbol (+) nex

Seite 38 - Data Viewing Window

Using the Single Module Data Viewer Software • 43month appear in the left side of the window. Expand the folder you want to see and click on the cali

Seite 39

Using the Single Module Data Viewer Software • 44the data view window that you wish to view. The contents in the file will appear in the lower right

Seite 40 - Viewing the Data

Using the Single Module Data Viewer Software • 45accidental deletion of instruments or data. To delete an instrument or data, perform the following:

Seite 41

Spare Parts List • 46 2. Click the OK button to exit the program or the Cancel button to return to the program.Spare Parts ListTable 6: Spare Parts L

Seite 42

Overview • 5 Chapter 1: IntroductionOverviewThis chapter briefly describes the SM-2001U and SM-2003U Single Module Calibration Stations and the Singl

Seite 43

6 • System RequirementsCAUTION: The GX-2001 and GX-2003 detect oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, a

Seite 44 - Deleting Data

Specifications • 7SpecificationsTable 1: SM-2001U & SM-2003U SpecificationsInput Power 100 VAC - 240 VAC, 50/60 Hz, 10 VANOTE: The power cable

Seite 45 - Exiting the Program

8 • About this ManualAbout this ManualThe SM-2001U & SM-2003U Single Module Calibration Stations Operator’s Manual uses the following conventions

Seite 46 - Spare Parts List

Overview • 9Chapter 2: DescriptionOverviewThis section describes the SM-2001U and SM-2003U single module calibration stations. They are designed to b