www.rkiinstruments.com GX-2009 Data LoggerManagement ProgramOperator’s ManualPart Number: 71-0163RKRevision: CReleased: 3/19/12
10 • IrDA Downloading CableWindows® Wireless Link Operation NoteWhen using an IrDA adapter cable and the GX-2009 Data Logger Management Program on a
IrDA Downloading Cable • 11 6. Click OK. 7. Close the Control Panel window. Figure 2: Image Transfer TabDeselect
12 • Launching the ProgramLaunching the Program 1. Click Start on the Windows® Icon Tray, then select All Programs, then select GX-2009. Your operati
Control Buttons • 13Control ButtonsThis section provides an overview of the control buttons. Instructions for using the various parts of the program
14 • Control ButtonsDownload ButtonIf you are in another program window, clicking the Download button opens the Download Window (see Figure 3). The D
Control Buttons • 15commands and is currently connected to the program. If an instrument is turned off after being connected to the program, then the
16 • Control ButtonsNOTE: If the GX-2009 Data Logger Management Program database is on the user’s computer (default setting), data will only appear f
Control Buttons • 17• Alarm Trend Data FilesAlarm trend data is logged around an alarm event. The GX-2009 can save up to 8 alarm trend data files in
18 • Control ButtonsThe Last Calibration Window stores the information for the most recent successful calibration and for the most recent successful
Control Buttons • 19Set ButtonClicking the Set button opens the Set Window.In the Set Window, you can perform the following functions:• Alter the app
2 • WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performan
20 • Accessing a Common DatabaseAccessing a Common DatabaseThe GX-2009 Data Logger Management Program stores downloaded data in a database file. This
Accessing a Common Database • 21last line. This line is the Data Location path. An example of the path created by the program is shown in Figure 9. R
22 • Accessing a Common Database 4. Save the file and reopen the GX-2009 Data Logger Management Program. 5. When you launch the program again, a da
Downloading Data from the GX-2009 • 23Downloading Data from the GX-2009You have the option of downloading data manually or automatically. If you want
24 • Downloading Data from the GX-2009IrDA adapter cable as shown in Figure 12 below, aligning the infrared port on the side of the GX-2009 with the
Downloading Data from the GX-2009 • 25Commands become selectable. 5. If you are going to download data manually, you can perform a complete download,
26 • Downloading Data from the GX-2009 7. If the unit has been downloaded manually (Automatic Download not selected), the unit’s instrument informati
Downloading Data from the GX-2009 • 27 8. You can view, print, export, or delete downloaded data by entering the Data or Last Calibration Windows. Th
28 • GX-2009 Data Logging CapacityGX-2009 Data Logging CapacityTable 1 above lists the GX-2009’s data logging capacity for each interval trend time
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 29Viewing, Printing, Exporting, and Deleting Data in the Data WindowThe GX-2009
Table of Contents • 3Table of ContentsWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
30 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowThe Data Window is divided into four frames. The upper left frame is the Data
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 31NOTE: In the examples that follow, the data will be shown organized by serial
32 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowinformation that can be accessed by the Last Calibration button.All downloaded
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 33Data Window’s upper left frame to see the folders of downloaded data. 3. Find
34 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 7. To view the Bump Test file in table format, double click the Bump Test fil
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 35type, and file location. The default file type is “.csv” (comma-separated valu
36 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowInstrument calibration information is also available in the Last Calibration W
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 37To view, print, or export the calibration history for any instrument in the da
38 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 4. Click the expanded view symbol (+) of or double-click the untitled folder.
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 39 7. To view the Calibration Histories file in table format, double click the C
4 • Table of ContentsInterval Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Alarm Trend Data . . . . . .
40 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowAs” dialog box will appear for you to specify the filename, file type, and fil
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 41instrument and save them in alarm event files and trouble event files for each
42 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowdownloaded data. 3. Find your instrument by serial number, then click the expa
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 43 7. Click the View Data button at the bottom of the Data Window or double-clic
44 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 9. If you want to print the data, click the Print button. A Printer List dial
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 45you want to view.Interval Trend DataAverage gas concentrations over the user d
46 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowTo view and perform desired operations with the interval trend files: 1. With
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 47dated folder whose contents you want to see. 5. Click on the Interval Trend Fo
48 • Viewing, Printing, Exporting, and Deleting Data in the Data Window• In table format, the interval trend data is shown as the average gas reading
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 49POWER MODE button on the GX-2009.• If you click the Summary button, the Data W
Introduction • 5IntroductionNOTE: The bump test function in Calibration Mode is included in instruments with firmware version 03955 or later installe
50 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 9. To view only events in the interval trend data file, click the Events Only
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 51• For any event, the data point before and after the event is always shown. 11
52 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowWhen viewing interval trend data in graph format, you have several options: •
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 53 12.To print the data, whether it is viewed in table or graph format, click th
54 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowFor Graph view, the export file type is Windows bitmap (.bmp). For Table view,
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 55To view and perform desired operations with the alarm trend files: 1. With the
56 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowthe upper right frame of the Data window, a list of file names will appear in
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 57• In table format, the log times are shown along with the peak (minimum for ox
58 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 8. To view the data in graph format, click the Graph button. An alarm trend f
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 59the gas reading and log time for each data point in the alarm trend file. As y
6 • System Requirements• view, print, and save data• change data logging parameters• change the appearance of the software screens• change the color
60 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 11.To export the data to another application, for example a spreadsheet or da
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 61• Interval Trend and Alarm Trend data filesTo delete any of the above items in
62 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 5. Enter the password and click Continue. The password when the program is fi
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 63Changing the PasswordCAUTION: Changing the password requires use of the Delete
64 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 3. Click Change Password. The Password Window asks you to input the current p
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 65 7. Click OK when the program confirms that you have changed the password. 8.
66 • Viewing, Printing, and Deleting Data in the Last Calibration WindowViewing, Printing, and Deleting Data in the Last Calibration WindowYou can ac
Viewing, Printing, and Deleting Data in the Last Calibration Window • 67Effects of Bump Test Function Parameter Setting on Appearance of Last Calibra
68 • Viewing, Printing, and Deleting Data in the Last Calibration WindowIf any instrument has been downloaded to the GX-2009 Data Logger Management P
Viewing, Printing, and Deleting Data in the Last Calibration Window • 69Operator’s Manual.Viewing and Printing Last Calibration DataOpen the Last Cal
Installing the GX-2009 Data Logger Management Program • 7• CD-ROM Drive• Infrared port orUSB port and a USB/IrDA adapter cableInstalling the GX-2009
70 • Viewing, Printing, and Deleting Data in the Last Calibration WindowThe Need Calibration/Need Bump Test view option shows the following fields fo
Viewing, Printing, and Deleting Data in the Last Calibration Window • 71NOTE: Your GX-2009 may have less than 4 sensors installed.• Last Downloaded —
72 • Viewing, Printing, and Deleting Data in the Last Calibration WindowCalibration Date/Bump Test Date View OptionSelecting this option shows the la
Viewing, Printing, and Deleting Data in the Last Calibration Window • 73To print a list of the instruments shown in the Calibration Date/Bump Test Da
74 • Viewing, Printing, and Deleting Data in the Last Calibration WindowThe fields on this screen include SerialNo, UserID, and StationID just as in
Viewing, Printing, and Deleting Data in the Last Calibration Window • 75It is not possible to print any information when the Calibration Record/Bump
76 • Changing GX-2009 Instrument Parameters 6. Enter the password and click the Continue button. The password when the software is first installed is
Changing GX-2009 Instrument Parameters • 77Instrument Information, and Power Off download commands will be selectable. Click the Instrument Informati
78 • Changing GX-2009 Instrument Parameterschange (e.g. H2S AutoCal.) to select it, then type the new information. 9. After you have finished enteri
Changing the Appearance of the Program Screens • 79Changing the Appearance of the Program ScreensTo change how information is displayed in the progra
8 • Installing the GX-2009 Data Logger Management Program 4. After a few seconds, a screen appears indicating that the InstallShield Wizard is prepar
80 • Changing the Appearance of the Program Screens 4. Select the colors used to graph the various target gases in the interval trend and alarm trend
Spare Parts List • 81• If you want to define a custom color, click the Define Custom Colors button and the Custom Color Window shown on the right abo
IrDA Downloading Cable • 9IrDA Downloading CableThe GX-2009 communicates with a computer via an on-board infrared communication port that complies wi
Kommentare zu diesen Handbüchern