59 Using Setup Mode GX-2003 Operator’s Manual
2. Use the AIR or (SHIFT) button to display the desired setting.
Tip: Press and hold the AIR or (SHIFT) button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Calibration Time Remaining Setting
With CAL. TIME REMAINING set to ON (factory setting), the GX-
2003 will give an indication at start up if it is due for calibration. The
type of indication will depend on the CAL. EXPIRED ACTION
setting (see below).
1. From the main menu, place the cursor next to the CAL. TIME
REMAINING menu item. Press and release the POWER ENTER
button.
2. Use the AIR or (SHIFT) button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Calibration Expired Action Setting
This item defines what indication is given during start up when
calibration is due and CAL. TIME REMAINING is set to ON.
With CAL. EXPIRED ACTION set to CONFIRM TO USE (factory
setting), the GX-2003 will give an indication at start up if calibration is
due and require the user to press the RESET SILENCE button to
continue.
With CAL. EXPIRED ACTION set to CANNOT USE, if the unit is
due for calibration the GX-2003 will give an indication at start up that
calibration is due and the unit cannot be used until it is calibrated.
With this setting, it is necessary to calibrate the instrument from the
Setup Mode if calibration is not performed before it is due. See
“Calibrating Using Auto Calibration” on page 61 and “Calibrating
Using Single Calibration” on page 63 for instructions.
With CAL. EXPIRED ACTION set to NO EFFECT, no confirmation
during startup or calibration is required to use the GX-2003.
CALIBRATION TIME
REMAINING
ON
21:09