www.rkiinstruments.com SDM-2009 Docking StationStandalone ConfigurationOperator’s ManualPart Number: 71-0261RKRevision: 0Released: 8/12/13

Seite 2 - Warranty

6 • USB CableUSB CableA Type A to Type B USB cable is provided with the docking station. It is only for use with the PC Controlled configuration. It

Seite 3 - Table of Contents

Air Filter and Sample Tubing • 7WARNING: Do not use an exhaust tube that is longer than 10 feet. The increased flow restriction caused by a longer tu

Seite 4

8 • Instrument CradleInstrument CradleThe instrument cradle is a recessed area on the top of the SM-2009 that is designed to accept the GX-2009. Inse

Seite 5 - Chapter 1: Introduction

Back Panel • 9Back PanelThe back panel includes the power jack, sample fittings, and a USB PC connector. Power JackThe power jack is located in the b

Seite 6 - System Requirements

10 • Control PanelControl PanelThe control panel is used to setup and operate the docking station in the Standalone configuration. It is located at t

Seite 7 - Specifications

Control Panel • 11CAL S • Initiates a calibration• Cancels a calibration• Clears data from docking station memory (when used with COPY button)• Move

Seite 8 - About this Manual

12 • Front PanelFront PanelA type A USB port is located on the front of the docking station. This port can be used to save calibration and bump test

Seite 9 - Chapter 2: Description

Overview • 13Chapter 3: Preparing to Use the SDM-2009OverviewThere are four tasks that must be completed before you can begin to use the SDM-2009: ha

Seite 10 - Air Filter and Sample Tubing

14 • Setting the Operational Parameters in Edit ModeSetting the Operational Parameters in Edit ModeOnce the hardware has been assembled, use Edit Mod

Seite 11 - For GAS

Setting the Operational Parameters in Edit Mode • 15Air Sample Time (AIr)The air sample time can be set separately for bump testing and calibration.

Seite 12 - Instrument Cradle

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Seite 13 - Back Panel

16 • Setting the Operational Parameters in Edit ModeTurning on the SDM-2009 with an InstrumentDo the following to turn on the SDM-2009 and establish

Seite 14 - SDM-2009

Setting the Operational Parameters in Edit Mode • 17screen shown above in step 6. The CHARGE LED will begin to blink green.• When Cal Limit Display i

Seite 15

18 • Setting the Operational Parameters in Edit Mode 4. If you wish to cancel setting the bump test parameters or were just viewing the parameters to

Seite 16 - Front Panel

Setting the Operational Parameters in Edit Mode • 19 9. When you are done setting the parameters, use the BUMP T button to scroll past the AUTO--C pa

Seite 17 - Hardware Assembly

20 • Setting the Operational Parameters in Edit ModeParameters” on page 14. 5. Use the BUMP T and CAL S buttons to scroll through the different cal

Seite 18

Connecting Calibration Gas • 21Connecting Calibration Gas The GAS fitting on the back of the docking station is designed to be used with a calibratio

Seite 19

22 • Installing the Single Module Data Viewer SoftwareInstalling the Single Module Data Viewer Software 1. Launch Windows®. 2. Exit from all applicat

Seite 20 - 50 25.0

Overview • 23Chapter 4: OperationOverviewWhen you have completed the tasks in "Chapter 3: Preparing to Use the SDM-2009", you are ready to

Seite 21

24 • Bump Testing a GX-2009and the SDM-2009 occurs, the home screen will appear on the instrument display at the end of the startup sequence. The aut

Seite 22

Bump Testing a GX-2009 • 25 9. Verify that the appropriate calibration gas cylinder is connected to the GAS fitting on the back of the SDM-2009. See

Seite 23 - BUMP/CAL

Table of ContentsTable of ContentsChapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Seite 24

26 • Bump Testing a GX-2009• If one or more of the sensors fails the fresh air adjustment, the SDM-2009 will abort the bump test and will not apply c

Seite 25 - Connecting Calibration Gas

Bump Testing a GX-2009 • 27for a description of calibration results. 16. The SDM-2009 will then purge the system with fresh air for the time defined

Seite 26 - Software

28 • Bump Testing a GX-2009 Figure 15: Screen Indication for Failed Bump Test and Failed Calibration 18. After a successful or failed bump test,• To

Seite 27 - Chapter 4: Operation

Calibrating a GX-2009 • 29a USB flash drive. 20. Remove the GX-2009 from the SDM-2009. 21. If you wish to bump test additional instruments, repeat s

Seite 28

30 • Calibrating a GX-2009NOTE: The screen shown above applies to a 4-channel GX-2009. If your GX-2009 has less than 4 channels, the inactive channel

Seite 29 - - - - - -

Calibrating a GX-2009 • 31 9. Press and hold the CAL S button until the CAL S LED turns on (about one second) then release it. During the calibration

Seite 30 - B/CAL G 1

32 • Calibrating a GX-2009 13. The SDM-2009 will apply calibration gas to the instrument for the time defined by the GAS calibration parameter. 14

Seite 31

Calibrating a GX-2009 • 33hold the EDIT ENTER and POWER buttons simultaneously for about one second and then release them. To avoid accidentally ente

Seite 32 - 24 13.0

34 • TroubleshootingTroubleshootingNOTE: This troubleshooting guide describes SDM-2009 problems only. See the GX-2009 Operator’s Manual for problems

Seite 33

Troubleshooting • 35• No IR connection between instrument and SDM-2009• SDM-2009 is not turned on.• The instrument is not correctly inserted into the

Seite 34 - C--LIMIT

Table of ContentsInstalling the Single Module Data Viewer Software. . . . . . . . . . . . . . . . . . . . . . . 22Chapter 4: Operation. . . . . . . .

Seite 35 - FLOW NG

36 • Charging an Instrument in a Docking StationCharging an Instrument in a Docking StationThe SDM-2009 can be used to charge the rechargeable batter

Seite 36 - CAL G 1

Charging an Instrument in a Docking Station • 37Recharging the Batteries Without Performing Any Operations 1. Confirm that the AC adapter is connecte

Seite 37

38 • Calibration and Bump Test RecordsCalibration and Bump Test RecordsThe SDM-2009 saves a record of each bump test and calibration performed. It is

Seite 38 - Troubleshooting

Calibration and Bump Test Records • 39SDM-2009 will take a few seconds to determine how much memory is available in the flash drive. • If the flash d

Seite 39

40 • Calibration and Bump Test RecordsThe files that an SDM-2009 saves to a USB flash drive are structured so that they can be imported into a databa

Seite 40

Overview • 41Chapter 5: Single Module Data Viewer ProgramOverviewThe Single Module Data Viewer Program is used to view, organize, and print bump test

Seite 41 - 04.59

42 • Using the Single Module Data Viewer ProgramNOTE: When you start the Single Module Data Viewer Program for the first time, there will be no data

Seite 42

Using the Single Module Data Viewer Program • 43• View the bump test and calibration data saved in the database.• Delete data.• Print bump test or ca

Seite 43

44 • Using the Single Module Data Viewer ProgramOrganizing the Data When viewing the data, it can be organized in two ways: 1. Base View FormatIn bas

Seite 44 - Multiple Times In One Day

Using the Single Module Data Viewer Program • 45 2. ID View FormatIn ID view format, the data can be organized by one or more of the following items

Seite 45

Overview • 1Chapter 1: IntroductionOverviewThis chapter briefly describes the SDM-2009 Docking Station and the Single Module Data Viewer Program. Thi

Seite 46

46 • Using the Single Module Data Viewer Program 2. When an item no longer has a (+) or (-) symbol next to it, single click it and the contents of th

Seite 47

Using the Single Module Data Viewer Program • 47If you organize the data by date, then folders organized by year/month appear in the left side of the

Seite 48 - Organizing the Data

48 • Using the Single Module Data Viewer ProgramIf you are viewing data in ID view format, expand the folders in the left side of the window until th

Seite 49 - Viewing the Data

Using the Single Module Data Viewer Program • 49appear showing the selected printer and confirming if you want to print. Verify that the selected pri

Seite 50

50 • Using the Single Module Data Viewer Program 3. Enter the password and click OK. The factory set password is “ABCDE” and is case sensitive. A con

Seite 51

Spare Parts List • 51 4. Enter the new password and click the New Password button. A confirmation window will appear prompting you for the new passwo

Seite 52

Spare Parts List • 5249-2058RK-04 4-port AC adapter71-0261RK Operator’s Manual, SDM-2009 Docking Station Standalone Configuration (this document)71-0

Seite 53 - Deleting Data

2 • System RequirementsCAUTION: The GX-2009 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Seite 54 - Changing the Password

Specifications • 3SpecificationsTable 1: SDM-2009 SpecificationsInput Power 12 VDCNOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz, 0.6A input and 12 VD

Seite 55 - Spare Parts List

4 • About this ManualAbout this ManualThe SDM-2009 Docking Station Standalone Configuration Operator’s Manual uses the following conventions for note

Seite 56

Overview • 5Chapter 2: DescriptionOverviewThis section describes the SDM-2009 docking station. It is designed to be used on a table top and consists