17 • Setting the Operational Parameters in Edit Mode
Setting the Bump Test Parameters
Do the following to set the bump test parameters after establishing a connection
between an EAGLE 2 and the SDM-E2.
1. Turn on the SDM-E2 with an EAGLE 2 and establish a connection between
them as described above in "Turning on the SDM-E2 with an EAGLE 2". The
EAGLE 2 will display the home screen.
2. Press and hold the EDIT ENTER button for 2 seconds. The EAGLE 2 will
display the Edit Mode screen.
3. Press and release the BUMP T button. The EAGLE 2 will display the following
screen with the four bump test parameters and their settings.
CH4 50%LEL GAS1
OXY 12.0vol% GAS1
H2S 25.0ppm GAS1
CO 50ppm GAS1
---
---
EDIT MODE
BUMP TEST : BUMP
CALIBRATION : CAL
GAS INLET : EDIT
BUMP TEST PARAMETER
1.AIR TIME: 30sec
2.GAS TIME: AUTO
3.CHECK : +-50%
4.AUTO CAL: OFF