EAGLE 2 Operator’s Manual Start Up, Leak Check Mode 283
12. The Date/Time Screen appears for a few seconds.
13. If USER/STATION ID is turned on (see “Turning the User/Station ID
Function On or Off” on page 118), the ID Screen appears for a few
seconds.
If USER/STATION ID is turned off, only the serial number is shown.
14. If the EAGLE 2 experiences a catalytic combustible sensor failure during
start up, a screen indicating which sensor failed appears and the buzzer
sounds a pulsing tone twice per second. In the example below, the H
2
S
sensor has failed.
If the catalytic combustible sensor fails, it is not possible to enter
Leak Check Mode. Press and release the POWER ENTER RESET
button to acknowledge the failure and return to the Mode Select
Screen. Replace the failed sensor.
If any other sensor that is installed fails, press and release the
POWER ENTER RESET button to acknowledge the failure and
continue to Leak Check Mode. Change the failed sensor(s) as soon
as possible for use in Normal Mode.
9/12/2008
15:00:00
USER ID
MIKE
STATION ID
PUMP 1
SERIAL NUMBER
E2A515
FAILED SENSOR(S)
<CH4> < >
< > < >
ENTER TO CONTINUE