Rki-instruments GX-2012 Manual Bedienungsanleitung Seite 30

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 185
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 29
22 Start Up, Normal Mode GX-2012 Operator’s Manual
If the unit is due for a bump test and BP EXPRD is set to NOT USE,
then the following screen displays.
The GX-2012 cannot be used until a bump test has been performed
by selecting BUMP in the Maintenance Mode menu. See
“Performing a Bump Test” on page 111 for bump testing instructions.
If a bump test is not due or BP EXPRD is set to NO EFFECT, then
the following screen appears for a few seconds indicating when the
next bump test is due.
NOTE: If both CL RMNDR and BP RMNDR are turned off, a WARM UP
screen will display before the warm up sequence continues.
FAIL
B--LIMIT
2011
9
30
10d
NEXTBUMP
Seitenansicht 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 184 185

Kommentare zu diesen Handbüchern

Keine Kommentare