Rki-instruments GX-94 Bedienungsanleitung Seite 25

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 52
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 24
GX-94 Operator’s Manual Page 24
Short-Term Exposure Limit (STEL) Screen
The STEL function shows the average reading for H
2
S and CO during
the last 15 minutes. These values are not affected by the Lunch Break
function and are reset every time the unit is turned off.
Time-Weighted Average (TWA) Screen
The TWA function shows the average reading for toxic gases during the
last 8 hours. If 8 hours has not elapsed since the last time the STEL/
TWA was cleared, the average is still calculated over 8 hours, with the
missing time assigned a 0 value for readings.
When the Lunch Break RESUME option is selected during start-up, the
GX-94 will include the TWA from the last session.
STEL
H2S 0.
0
PPM
CO 0PPM
TWA
H2S 0.
0
PPM
CO 0PPM
Seitenansicht 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 51 52

Kommentare zu diesen Handbüchern

Keine Kommentare