30-0954RK-200 Series Sample Draw Aspirator Adapter • 4
The drain fault switch senses a break or very low flow in the water drain line and provides open
contacts in normal operation in the low flow/drain fault contact housing which close when the
switch senses a break, a very low drain flow, or a shut down.
Connecting monitoring devices to the low flow and drain fault contact terminals provides the user
with a notification of a contact closure. Each set of contacts may have their own monitoring device
or they may share one. See “Installation” on page 4 for instructions to wire these terminals using
the two scenarios.
A second plastic cable bushing on the bottom of the low flow/drain fault contact housing allows for
cable entry to the housing. The size range of the cable which can be routed through the second
cable bushing is .064’’ - .210’’ OD.
Installation
Figure 3: Outline & Mounting Dimensions
1. Install the mounting panel to a flat vertical surface using the four mounting holes (0.50’’
diameter) in the corners of the panel. Make sure that sufficient space is allowed to remove the
Low Flow and Drain Fault
Contact Housing
IN
Field Wiring Cable
Entry Bushing for
Switch Contact
Sample Inlet
Aspirator/Sample Exhaust
0.75
Detector
Head with
Chamber
Detector
Head with
Chamber
(4X) Ø 0.50
OUT
1
IN
OUT
2
(optional)
140 PSI Compressed
Air Inlet
+
-
0.75
19.25
20.75
19.25
20.75