GX-2003 Operator’s Manual Calibrating the GX-2003 82
5. Slide the tubing clamp onto the tubing and connect the tubing
onto the sample bag’s inlet. Leave the clamp unclamped for now.
6. Connect the other end of the tubing to the regulator’s or
dispensing valve’s hose barb fitting.
7. Fill the gas collection bag by screwing the fixed flow regulator
onto the calibration cylinder or by screwing the dispensing valve
onto the calibration cylinder and turning the knob
counterclockwise.
NOTE: If the regulator has no On/Off knob, then gas will begin to
flow at 0.5 LPM as soon as the regulator is connected to the
cylinder.
8. Allow the gas to dispense until the gas collection bag is a little over
half full.
9. If a fixed flow regulator without a knob is used, disconnect it from
the cylinder and immediately clamp down the hose clamp to
prevent the release of gas from the gas collection bag. If a fixed
flow regulator with a knob or a dispensing valve is used, turn the
knob clockwise to stop the gas flow, clamp down the hose clamp
and remove the regulator or dispensing valve from the cylinder.
10. Disconnect the tubing from the regulator or dispensing valve.
Calibrating the GX-2003
From Measuring Mode, enter Calibration Mode by pressing and
holding the (SHIFT) button, then pressing the DISPLAY(ADJ)
button and releasing both. If the unit prompts you for the password,
enter it by using the AIR and (SHIFT) buttons to select each
password number and then pressing and releasing POWER ENTER to
enter it and move on to the next one. The calibration menu displays
with the cursor next to AUTO CALIBRATION.
NOTE: The following screens illustrate a four-gas GX-2003 with both
a % LEL and a % volume combustible sensor as examples
only. Your GX-2003 may display slightly different screens.
>AUTO CALIBRATION
SINGLE CALIBRATION
CONTRAST
NORMAL OPERATION
21:09