
Bump Testing Standard 4-Gas Instrument(s) • 52
8. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the
bump test. When using a 3-gas (no H
2
S channel) or a 4-gas (H
2
S channel
included) cylinder, to bump test a standard instrument, select all channels. If
your instrument has fewer than 4 channels, your screen will only display the
active channels.
Figure 33: Gas Cylinder Confirmation Screen
When all of the desired channels are selected, click OK. If you don’t want to
proceed, click Cancel to abort the bump test.
Kommentare zu diesen Handbüchern