Calibrating an Instrument with Special Sensors 92
If you are calibrating an instrument or instruments with 2 special sensors, this
confirmation screen will indicate that the calibration cylinder for the first special
sensor (in this case, NH
3
) must be connected to the GAS 2 fitting first and that
you must change the GAS 2 cylinder midway through the calibration (in this
example, to IBL). The GAS 2 shown in the middle column will always be
calibrated before the GAS 2 shown in the right column. The GAS 2 selection
boxes are deselected by default. You must select the GAS 2 boxes if you wish
to calibrate the special sensors. If you do not wish to calibrate the special
sensors, leave the GAS 2 boxes deselected. Be sure that the appropriate GAS
2 cylinder is connected to the GAS 2 fitting.
Figure 69: Multiple GAS 2 Cylinder Confirmation
9. Click OK. You will be asked if you want to calibrate the instrument(s). Click OK
to proceed. If you don’t want to proceed with the calibration, click Cancel to
return to the main program window.